Singapore: HSA

Registration evidence and traceability for HSA submissions

For medical device teams registering devices in Singapore. Keep the design verification, risk analysis and supporting documents behind an HSA registration linked and controlled, so the dossier is assembled from approved, current documents.

See medical device featuresBook a demo

In Singapore, the Health Sciences Authority (HSA) regulates medical devices under the Health Products Act and the Health Products (Medical Devices) Regulations 2010. Most devices must be registered before they are supplied, with a dossier of design, risk and verification evidence. Traceable keeps that evidence controlled and traced, so the dossier draws on approved documents.

What an HSA dossier asks for, and where it comes from in Traceable

Documents HSA lists for a full-route registration dossier, and where each comes from in Traceable
Dossier documentWhere it comes from in Traceable
Essential principles checklistAn ID grid with one row per principle, each linked to the requirement, test or report that shows conformity.
Device descriptionA controlled document whose published version, with its signatures, is the one the dossier cites.
Design verification and validationRequirements traced to the tests that verify them in a live traceability matrix, with gaps and broken links flagged.
Risk analysisA hazard analysis built as an ID grid with a computed rating, linked to its risk controls and their verification.
Clinical evaluation report and labellingControlled documents with review, e-signatures and a locked PDF snapshot at each release.
Quality management system evidenceProcedures and records under document control, with an attributed change log behind every revision.

Risk classes, registration and dealer licences

HSA sorts medical devices into four risk classes, from Class A (lowest risk) to Class D (highest), according to the device's intended purpose, how it works, who uses it and the technology involved. Class A devices are exempt from product registration; most others must be registered before they can be supplied in Singapore. Companies that manufacture, import or wholesale devices also need a dealer's licence.

The registration route depends on the class and on whether the device already has approval from one of HSA's reference regulatory agencies. A full route asks for the most evidence; abridged, expedited and immediate routes ask for less where those approvals exist. Submissions use the ASEAN Common Submission Dossier Template (CSDT), and documents must be in English.

The guidance a registration is built on

HSA publishes numbered guidance documents that set out how it applies the regulations. For a device team, the ones that shape the documentation include:

  • GN-13 and GN-14: risk classification of general medical devices and of in vitro diagnostics.
  • GN-17 and GN-18: preparing a product registration submission using the ASEAN CSDT, for general medical devices and for in vitro diagnostics.
  • GN-05: reporting adverse events, a duty on every company dealing in devices, registrants included.
  • GL-04 and GL-07: software medical devices across their life cycle, and the risk classification of software as a medical device and qualification of clinical decision support software.

HSA revises these documents regularly. Work from the current version on HSA's guidance documents page, and check the registration guide for your class and route.

Design controls, traceability and document control in one place

Document control

Review gated on resolved comments, e-signatures with the signer, role and time, and a locked snapshot at every publish. Our ISO 13485 document control guide covers the quality system side.

Design controls

Lay out design inputs, outputs, verification, validation and transfer on the project plan, and keep the design history file current, so the verification and validation the dossier summarises already exist.

Traceability

Trace user needs to requirements to tests in a live traceability matrix. See our guide to the traceability matrix for medical devices.

Risk management

Build the risk analysis as an ID grid with severity, occurrence and a computed rating, linked to the controls and tests. Our ISO 14971 guide explains the process.

Software medical devices

Start from templates aligned with IEC 62304 for the software life cycle evidence. Read the IEC 62304 guide and use the free IEC 62304 checklist.

One set of documents, many markets

Branch or clone a document where a market needs its own version, and keep the shared design and risk evidence traced once. Begin from the free traceability matrix template.

What Traceable does not do

Traceable is documentation software. It does not submit registrations to HSA, hold a dealer's licence, act as your registrant, report adverse events or decide your device's risk class, and it is not approved or endorsed by HSA. It keeps the documents a registration depends on controlled and traced.

Teams that also sell into Europe or the UK can read the EU MDR and HPRA page and the MHRA and UKCA page.

Common questions

Is Traceable approved by HSA?

No. HSA registers medical devices and licenses the companies that deal in them; it does not approve the software a manufacturer uses for its documentation. Traceable helps you produce, control and trace the evidence a registration submission draws on. Registration remains the registrant’s responsibility.

Can we reuse our EU or US documentation for an HSA submission?

Often in part. HSA’s abridged, expedited and immediate routes take account of prior approval by its reference agencies, and the dossier follows the ASEAN Common Submission Dossier Template. In Traceable the underlying design, risk and verification documents stay the same, traced once, whichever market a dossier is for.

Does HSA regulate software as a medical device?

Yes. Software with a medical purpose can be a medical device in Singapore, and HSA publishes guidelines on software medical devices across their life cycle and on the risk classification of software. Traceable is not a medical device and makes no classification decision for you.

Does Traceable submit to HSA for us?

No. Submissions are made through HSA’s own systems by the registrant. Traceable keeps the documents behind the dossier current and approved, so the dossier is assembled from approved documents rather than chased together.

See your registration evidence traced

Import your existing requirements and risk files, and see live coverage in minutes.

Start free